TCP312 w/TCPA300 |
TCP305 w/TCPA300 |
TCP303 w/TCPA300 |
TCP404XL w/TCPA400 |
Bandwidth (-3 dB) |
DC ¨C 100 MHz |
DC ¨C 50 MHz |
DC ¨C 15 MHz |
DC ¨C 2 MHz *1 |
Risetime |
¡Ü3.5ns |
¡Ü7ns |
¡Ü23ns |
¡Ü175ns |
Maximum Current Ratings: |
High Current Sensitivity Range
DC (continuous)
RMS (sinusoidal)
Peak Pulse |
10 A/V Range
30A
21.2A
50A |
10 A/V Range
50A
35.4A
50A |
50 A/V Range
150A
150A
500A |
1A/mV Range
500A (750A *2)
500A
750A |
Low Current Sensitivity Range
DC (continuous)
RMS (sinusoidal)
Peak Pulse |
1 A/V Range
5A
3.5A
50A |
5 A/V Range
25A
17.7A
50A |
5 A/V Range
25A
17.7A
500A |
N/A |
DC Accuracy
(Operating temp 0 ¡ãC to 50 ¡ãC) |
¡À3% of reading |
¡À3% of reading |
¡À3% of reading (10 ¡ãC to 50 ¡ãC)
+3%/-6% of reading (0 ¡ãC to <10 ¡ãC) |
¡À3% of reading |
DC Accuracy, Typical
(Operating temp 23 ¡ãC ¡À5 ¡ãC) |
¡À1% of reading |
¡À1% of reading |
¡À1% of reading |
¡À1% of reading |
Nominal |
Maximum Bare Wire Voltage |
For Use With Insulated Wires Only |
600V RMS CAT I & II
300V RMS CAT III |
Lowest Measurable Current
(@ ¡À 3% accuracy at DC) |
1mA |
5mA |
5mA |
1A |
Scope Set To 1mV/div and 20 MHz BW Limited |
Scope Set To 1mV/div and 20 MHz BW Limited |
Insertion Impedance
(See Curves Below) |
0.08 ¦¸@ 1MHz
0.15 ¦¸@ 10MHz
0.27 ¦¸@ 50MHz
0.7 ¦¸@ 100MHz |
0.035 ¦¸@ 1MHz
0.12 ¦¸@ 10MHz
0.4 ¦¸@ 50MHz |
0.01 ¦¸@ 1MHz
0.025 ¦¸@ 5MHz
0.1 ¦¸15MHz |
0.1 m¦¸ @ 10kHz
0.6 m¦¸@ 100kHz
8 m¦¸ @ 1MHz
16 m¦¸ @ 2MHz |
Typical |
Maximum Amp-Second Product (Based on Amplifier Range setting) |
50A*¦ÌS ¨C 1A/V
500A*¦ÌS ¨C 10A/V |
500A*¦ÌS ¨C 5A/V
NA ¨C 10A/V |
3,000A*¦ÌS ¨C 5A/V
15,000A*¦ÌS ¨C 50A/V |
NA ¨C 1A/mV |
AC Coupling Low Frequency Bandwidth
(Low Pass - 3 dB point) |
< 7 Hz |
Displayed RMS Noise
(@ 20MHz Bandwidth Limit) |
¡Ü250 ¦ÌA RMS |
¡Ü1.25 mA RMS |
¡Ü2.5 mA RMS |
¡Ü250 mA RMS |
Signal Delay (Delay to Output BNC) |
17 ns |
19 ns |
53 ns |
103 ns |
Inputs (probe amplifier) |
1 |
Probe Open Indicator |
YES |
Overload Indicator |
YES |
Termination Indicator |
YES |
Non-Compatible Probe Indicator |
YES |
Safety Certifications |
U.S. NRTL Listing |
UL3111-1
(Amplifier) |
UL3111-2-032 ; UL3111-1
(Probe and Amplifier) |
Canadian Certification |
CAN/CSA C22.2 No.1010.1
(Amplifier) |
CAN/CSA C22.2 No.1010.2.032
CAN/CSA C22.2 No.1010.1
(Probe and Amplifier) |
European Union Compliance |
EN61010-1/A2
(Amplifier) |
EN61010-1/A2; EN61010-2-032
EN61010-1/A2
(Probe and Amplifier) |
Other |
IEC61010-1/A2 (Amplifier) |
IEC61010-2-032
IEC61010-1/A2
(Probe and Amplifier) |
Electromagnetic Compatibility |
EC Council Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS 2064.1/2. |
Power Requirements
(TCPA300/TCPA400 Amplifiers) |
90V to 264V; 47 to 440 Hz; 50 W
Maximum CAT II (Auto Switch) |
Power Requirements (Probes) |
Requires TCPA300 Amplifier |
Requires TCPA400 Amplifier |
Probe Model |
TCP312 |
TCP305 |
TCP303 |
TCP404XL |
Warranty |
1 Year |
Probe Mechanical Characteristics |
Probe Cable Length |
1.5 meters (60 inches) |
2 meters (78.7 inches) |
8 meters (315 inches) |
Probe Jaw Size (Max Conductor Size) |
3.8 mm (0.15 inches) |
21 mm x 25 mm (0.83 x 1.0 inches) |
Length:
Width:
Height:
Weight: |
20 cm (7.87 inches)
1.6 cm (0.625 inches)
3.2 cm (1.25 inches)
0.15 kg (0.33 lb) |
26.8 cm (10.55 inches)
4.1 cm (1.60 inches)
15.6 cm (6.13 inches)
0.66 kg (1.45 lb) |
26.8 cm (10.55 inches)
4.1 cm (1.60 inches)
15.6 cm (6.13 inches)
0.88 kg (1.90 lb) |
TCPA300 and TCPA400 Mechanical Characteristics |
Length:
Width:
Height:
Weight: |
17.3 cm (6.8 in)
16.7 cm (6.6 in)
9.14 cm (3.6 in)
1.14 kg (2.5 lb) |
Environmental Characteristics |
Temperature - Operating |
0 ¡ãC to +50 ¡ãC (32 ¡ãF to 122 ¡ãF) |
Temperature - Nonoperating |
−40 ¡ãC to +75 ¡ãC (−40 ¡ãF to 167 ¡ãF) |
Humidity ¨C Operating |
5% to 95% R.H.to +30 ¡ãC (86 ¡ãF)
5% to 85% R.H.+30 ¡ãC to +50 ¡ãC (86 ¡ãF to 122 ¡ãF) |
Humidity ¨C Nonoperating |
5% to 95% R.H.to +30 ¡ãC (86 ¡ãF)
5% to 85% R.H.+30 ¡ãC to +75 ¡ãC (86 ¡ãF to 167 ¡ãF) |
Altitude ¨C Operating |
2000 m (6800 ft) maximum |
Altitude ¨C Nonoperating |
12,192 m (40,000 ft) maximum |