Circuit method |
Bipolar sink and source |
Constant Voltage |
2.5 V: 0 to }2.5000 V (}0.03 % of setting }300 ƒÊV, 100 ƒÊV resolution) 25 V: 0 to }25.000 V (}0.03 % of setting }3 mV, 1 mV resolution) |
Constant Current |
25 mA: 0 to }25.000 mA (}0.03 % of setting }3 ƒÊA, 1 ƒÊA resolution) |
Thermoelectric power generation |
K: at TC: 0 ‹C, -174.0 to 1372.0 ‹C (}0.05 % of setting }0.5 ‹C, 0.1 ‹C resolution), Other types: E, J, T, R, S, B, N selectable |
Thermoelectric power generation |
K: at TC: RJ, -174.0 to 1372.0 ‹C (}0.05 % of setting }1.0 ‹C, 0.1 ‹C resolution), Other types: E, J, T, R, S, B, N selectable |
Standard resistance (Rs) |
100 Ħ (}0.2 Ħ) |
Automatic generation |
Number of memory steps: 20 Interval time: 1 to 99 sec (at CV, CC, TC mode) | |
|
[Measurement functions] |
Voltage |
2.5 V: 0 to }2.8000 V (}0.03 % rdg. }300 ƒÊV, 100 ƒÊV resolution, 1 Mƒ¶ input resistance) 25 V: 0 to }28.000 V (}0.03 % rdg. }3 mV, 1 mV resolution, 1 Mƒ¶ input resistance) |
Current |
25 mA: 0 to }28.000 mA (}0.03 % rdg. }3 ƒÊA, 1 ƒÊA resolution, 25 ƒ¶ input resistance) |
Temperature |
-25.0 to }80.0 ‹C (}0.5 ‹C at 23 }5 ‹C, 0.1 ‹C resolution, use with the RJ sensor 9184) |
Sampling rate |
Approx. 1.67 times/sec | |
|
Additional functions |
Zero adjustment, Overflow display, USB communication, Monitor |
Power supply |
AC ADAPTER 9445-02/03 (100 to 240 V AC 50/60 Hz, 9 VA), Ni-MH battery HR6 ~ 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery ~ 4, 6 VA |
Dimensions, mass |
104 mm (4.09 in) W ~ 180 mm (7.09 in) H ~ 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 ~ 4 batteries) |
Supplied accessories |
INPUT CORD 9168 ~ 1, TEST LEAD 9170-10 ~ 1, Fuse ~ 1, LR6 (AA) alkaline cell ~ 4, Operation manual ~ 1 | |
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
Improve stability and reduce calibration costs compared with the previous HIOKI model
For Instrumentation systems (4 - 20 mA) and loop testing
Check temperature control equipment and electric distribution
8 types of thermocouples to test thermoelectric power generation
Ideal for electrical device testing and routine maintenance of production equipment such as calibrators
Use the max. 25 mA DC sink as an electric load
|
|
|
|
|
|
|
|
|