Fluke 724 Calibrador de temperatura
Measurement Accuracy |
Voltage DC |
30.000 V |
0.02%+ 2 counts (upper display) |
20.000 V |
0.02%+ 2 counts (lower display) |
100.00 mV |
0.02%+ 2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) | |
Current DC |
24.000 mA |
0.02%+ 2 counts | |
Resistance |
0.0 to 400.0 目 |
0.1目 (4-wire), 0.15目 (2- and 3-wire) |
401 to 1500 目 |
0.5目 (4-wire), 1目 (2- and 3-wire) |
1500 to 3200 目 |
1目 (4-wire), 1.5目 (2- and 3-wire) | |
Source Accuracy |
Voltage DC |
100.00 mV |
0.02%+2 counts |
10.000 V |
0.02%+2 counts |
-10.00 mV to 75.00 mV |
0.025 % + 1 count (via TC connector) | |
Resistance |
15.0 to 400.0 目 |
0.15 目 (exc. current 0.15 to 0.5 mA), 0.1 目 (exc. current 0.5 to 2 mA) |
401 to 1500 目 |
0.5 目(excitation current 0.05 to 0.8 mA) |
1500 to 3200 目 |
1 目(excitation current 0.05 to 0.4 mA) | |
RTDs and Thermocouples |
Measure Accuracy: |
NI-120 |
0.2 ﹉C |
PT-100 (385) |
0.33 ﹉C |
PT-100 (393) |
0.3 ﹉C |
PT-100 (JIS) |
0.3 ﹉C |
PT-200 (385) |
0.2 ﹉C |
PT-500 (385) |
0.3 ﹉C |
PT-1000 (385) |
0.2 ﹉C |
Resolution: |
0.1 ﹉C |
J: |
0.7 ﹉C |
K: |
0.8 ﹉C |
T: |
0.8 ﹉C |
E: |
0.7 ﹉C |
R: |
1.8 ﹉C |
S: |
1.5 ﹉C |
B: |
1.4 ﹉C |
L: |
0.7 ﹉C |
U: |
0.75 ﹉C |
N: |
0.9﹉C |
Resolution |
J, K, T, E, L, N, U: 0.1 ﹉C, 0.1 ﹉F B, R, S: 1 ﹉C, 1 ﹉F |
XK |
0.6﹉C |
BP |
1.2﹉C | |
Source Accuracy: |
NI-120 |
0.2 ﹉C |
PT-100 (385) |
0.33 ﹉C |
PT-100 (393) |
0.3 ﹉C |
PT-100 (JIS) |
0.3 ﹉C |
PT-200 (385) |
0.2 ﹉C |
PT-500 (385) |
0.3 ﹉C |
PT-1000 (385) |
0.2 ﹉C |
Resolution: |
0.1 ﹉C |
Note: |
Accuracy stated for 4-wire measurement. |
J: |
0.7 ﹉C |
K: |
0.8 ﹉C |
T: |
0.8 ﹉C |
E: |
0.7 ﹉C |
R: |
1.4 ﹉C |
S: |
1.5 ﹉C |
B: |
1.4 ﹉C |
L: |
0.7 ﹉C |
U: |
0.75 ﹉C |
N: |
0.9 ﹉C |
Resolution |
J, K, T, E, L, N, U: 0.1 ﹉C, B, R, S: 1 ﹉C |
XK |
0.6﹉C |
BP |
1.2﹉C | |
Specifications |
Ramp Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Ramps |
Slow ramp, Fast ramp, 25% step-ramp | |
Loop Power Function |
Voltage: |
24V |
Accuracy: |
10% |
Maximum current: |
22 mA, short circuit protected | |
Step Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Steps |
25% of range, 100% of range | |
Environmental Specifications |
Operating Temperature |
|
Storage Temperature |
|
Humidity (Without Condensation) |
90% |
10足C to 30足C |
75% |
30足C to 40足C |
45% |
40足C to 50足C |
35% |
50足C to 55足C | |
Safety Specifications |
Safety Rating |
|
CSA C22.2 No. 1010.1:1992 | |
EMC |
|
EN50082-1:1992 and EN55022:1994 Class B | |
Mechanical & General Specifications |
Size |
|
Weight |
|
Batteries |
|
Four AA alkaline batteries. | |
Warranty |
|
Battery Life |
|
Shock & Vibration |
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
Fluke 724 Calibrador de temperatura
Built for the masters of temperature.
The new Fluke 724: everything you need to troubleshoot and calibrate temperature.
-
Now you can carry one tool to expertly test all the temperature sensors and transmitters in your plant.
-
The Fluke 724 is the master*s answer to temperature calibration. The Fluke 724 can measure and source 10 thermocouple types and 7 RTD types, plus volts and ohms.
-
The Fluke 724 even handles high-speed pulsed RTD circuits and provides loop power.
-
The dual line measure/source display lets you source temperature and view loop current at the same time.
-
With its simple, ※no menus§ controls, the Fluke 724 is easy to operate, too. |
Feel the Heat! The new Fluke 724 Temperature Calibrator is a powerful yet easy-to-use calibrator. Use the measure and source functions to test and calibrate almost any temperature instrument.
- Easy to read dual display lets you view input and output simultaneously
- Measure RTDs, thermocouples, ohms, and volts to test sensors and transmitters
- Source/simulate thermocouples, RTDs, volts, and ohms to calibrate transmitters
- Perform fast linearity tests with 25% and 100% steps
- Execute remote tests with auto step and auto ramp
- Power transmitters during test using loop power supply with simultaneous mA measurement
- Store frequently-used test setups for later use
- Compact, streamlined shape makes it easy to carry
- Rugged, reliable design stands up to field use
- Backlight lets you work in poor light
- Large battery capacity of four AA cells
- Battery door for easy changes
|
|
|
|
|
|
|
|
|
|