Input Specifications
Voltage
115/230 VAC ± 10%, user selectable
Frequency
50/60 Hz ± 5%
Fuse
3.15 A, fast acting 250 VAC
Dielectric Withstand Test Mode
Output Rating
5000 V @ 20 mA AC
6000 V @ 7.5 mA DC
Voltage Setting
Range: 0 - 5.00 KV AC
0 - 6.00 KV DC
Resolution: 0.01 KV
Accuracy: ± (2% of setting + 5 V)
Maximum Limit
AC Range: 0.00 - 20.00 mA
Resolution: 0.01 mA
DC Range: 0 - 7500 ìA
Resolution: 1 ìA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Minimum Limit
AC Range: 0.000 - 9.999 mA
Resolution: 0.001 mA
DC Range: 0.0 - 999.9 ìA
Resolution: 0.1 ìA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Arc Detection
Range: 0 - 9, 0 disabled
Ground Fault Interrupt
GFI Trip Current: 450 ìA max (AC or DC)
HV Shut Down Speed: < 1ms
Current Display
Auto Range
AC Range 1: 0.000 - 3.500 mA
Range 2: 3.00 - 20.00 mA
DC Range 1: 0.0 ìA - 350.0 ìA
Range 2: 0.300 mA - 3.500 mA
Range 2: 3.00 mA - 7.50 mA
Accuracy All Ranges ± (2% of reading + 2 counts)
DC Output Ripple
² 5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load
Discharge Time
² 200 ms
The maximum capacitive load vs output voltage:
0.20 ìF < 1 KV 0.050 ìF < 4 KV
0.10 ìF < 2 KV 0.040 ìF < 5 KV
0.06 ìF < 3 KV 0.015 ìF < 6 KV
AC Voltage Waveform
Sine Wave, Crest Factor = 1.3 - 1.5
Output Frequency
Range: 50 or 60 HZ, User Selectable
Output Voltage Regulation
± (1% of output + 5 V) from no load to full load and over
input voltage range.
Dwell Timer
Range: AC 0, 0.3 - 999.9 sec (0 = Constant)
DC 0, 0.4 - 999.9 sec (0 = Constant)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ramp Timer
Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec
DC 1.0 - 999.9 sec (0=OFF)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Continuity Current
DC 0.1 A ± 0.01 A, fixed
Ground Continuity
Range: 0.0 ½ - 1.50 ½
Maximum Limit Resolution: 0.01
½
Minimum Limit Accuracy: ± (3% of setting + 0.02
½)
Ground Continuity
Range: 0.0 ½ - 0.50 ½
Auto Offset Resolution: 0.01
½
Accuracy: ± (3% of setting + 0.02
½)
Insulation Resistance Test Mode
Voltage Setting
Range: 30 - 1000 VDC
Accuracy: ± (2% of setting + 5 V)
Resistance Display
Range: 1 - 9999 M½ (4 Digit, Auto Ranging)
Resolution: 500 VDC - 1000 VDC
M
½ M½
0.001 1.000 - 9.999
0.01 10.00 - 99.99
0.1 100.0 - 999.9
1 1000 - 9999
Accuracy: ± (2% of reading + 2 counts) at test voltage
500 - 1000 V and 1 - 999.9 M
½
± (5% of reading + 2 counts) at test voltage
500 - 1000 V and 1000 - 9999 M
½
± (8% of reading + 2 counts) at test voltage
30 - 500 V and 1 - 1000 M
½
Maximum Limit
Range: 0, 1 - 9999 M½ (0=OFF)
Accuracy: Same as Resistance Display
Minimum Limit
Range: 1 - 9999 M½
Accuracy: Same as Resistance Display
Ramp Timer
Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 1.0 - 999.9 sec (0=OFF)
Accuracy: ± (0.1% of reading + 0.05 sec)
Delay Timer
Range: 0, 0.5 - 999.9 sec (0 = Constant)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Fault Interrupt
GFI Trip Current: 450 ìA max
HV Shut Down Speed: < 1 ms
General Specifications
Interface
RS-232 interface standard for entry-level automation.
Remote Control &
The following input and output signals are provided
Signal Output
through two 9 pin D type connectors:
1. Remote control: Test, Reset, and Remote Interlock
2. Remote recall of memory program #1, #2, and #3
3. Outputs: Pass, Fail, Test-in-process, and Reset
Program Memory
10 Memories, 3 steps per memory.
Security
Key Lock capability to avoid unauthorized access to
all test parameters and memory locations.
Mechanical
Bench or rack mount with tilt up feet.
Dimensions
(W x H x D) 8.46 x 3.5 x 14.57 in.
(215 x 89 x 370 mm)
Weight
20.96 lbs (9.53 kgs)
Calibration
Traceable to National Institute of Standards &
Technology (NIST). Calibration controlled by software.
Adjustments are made through front panel keypad in
a restricted access calibration mode. Calibration
information stored in non-volatile memory.
Available Accessories
Probe
(38081) High voltage retractable 5 KV AC probe.
Return Probe
(38082) Retractable probe used on the return side of the Hypot III.
Safe-T-Probe
(38814) Test Gun with trigger that controls retractable probe and
activates the instrument’s high voltage circuit.
Footswitch
(35822) Interlocked footswitch that provides a “hands-off”
remote start.