Constant Resistance Mode - Program : CR Resolution*2 - 1/16000 Of Rated Value | *1 All Mode Specification measure by 25˚C room temperature unless otherwise specified *2 Transient Mode Specification must be x2 |
Common | Constant Voltage Mode | (0-V) V | CVMedium Range | 0 - ( V / 2) V | CVLow Range | 0 - ( / 10) V | Temp Coefficient | 100 ppm / °C of Rated Voltage | Transient Time Range | Fast Band(default, Osc1) | 0.500 ~ 51.19 ms | Slow Band(Osc2, Osc3) | 0.500 ~ 511.9 ms | CV Resolutions*2 | 1/16000 of rated voltage | CV Accuracy*2 (CVH, CVM, CVL) | 0.05% +/- (0.1% x Vmax) V | Display Specifications | CV Resolution | 1/16000 of Rated Voltage | CV Accuracy (CVH, CVM, CCL) | 0.05% +/- (0.1% x V) V |
Constant Power Mode | CPHigh Range | (0-P) W | CPMedium Range | 0 - (P/2) W @ DC input current ≤ (I/2) A | CPLow Range | 0 - (P/10) W @ DC input current ≤ (I/10) A | Transient Time Range | Same as CC Mode | Temperature Coefficient | 300 ppm / °C of Rated Power |
Constant Power Mode : Program | CPHigh Accuracy*2 | 1.00% +/- (Px0.5%) W @ input current > (I/20) A, input voltage > (V/10) V | CPMedium Range | 1.00% +/- (Px0.5%) W @ input current > (I/100) A, input voltage > (V/10) V | CPLow Range | 1.00% +/- (Px0.5%) W @ input current > (I/1000) A, input voltage > (V/5) V | Program | CP Resolution*2 1/16000 of Rated Power |
Constant Current Mode | CCHigh Range | 0 - I A | CCMedium Range | 0 - (I/2) A | CCLow Range | 0 - (I/10) A | Transient Time Range | Fast Band (default, Osc1) | 0.050 ~ 51.19 ms | Slow Band (Osc2, Osc3) | 0.500 ~ 511.9 ms | Temperature Coefficient | 100 ppm / °C of Rated Current |
Constant Current Mode : Program | CC Resolution*2 | 1/16000 of rated current | CCHigh Accuracy*2 LHM | 0.05% +/- (Ix0.1%) A |
Constant Resistance Mode | Transient Time Range : CRM / CRH | Same As CC Mode | Transient Time Range : CRL | Same As CV Mode | Temperature Coefficient : CRM / H | 300 ppm / °C of Minimum Resistance | Temperature Coefficient : CRL | 300 ppm / °C of Maximum Resistance |
Constant Resistance Mode - Program | CR Resolution*2 | 1/16000 of rated value |
External Programming Mode | Monitor Output Signal | 0-10 Volts output for 0 to full scale value | VMON Accuracy | 0.10% +/- (Vx0.1%) V | IMON Accuracy | 0.10% +/- (Ix0.1%) A | Analog Program | 0~10 Volts Input yields 0 -- selected full scale loading in all modes | Accuracy | Same As Internal ± 0.1% Rating | Input Impedance | 200 k Ω ± 1 % | BandWidth(-3dB) | Limited By Internal Transient Time | Remote Interface | GPIB / RS-232 / ETHERNET / USB |
Programmable Protection | Power (OPP) | Range | (Px1.05/800) ~ (Px1.05) W | Resolution | (Px1.05/8000) W | Accuracy | 0.50% +/- (P x 21 / 8000) W | Voltage (OVP) | Range | (Vx1.05/1600) ~ (Vx1.05) V | Resolution | (Vx1.05/16000) V | Accuracy | 0.20% +/- (Vx1.05/800) V | Current (OCP) | Range | Ix1.05/1600) ~ (Ix1.05) A | Resolution | (Ix1.05/16000) A | Accuracy | 0.20% +/- (Ix1.05/800) A | Under Voltage Lockout (UVL) | Mode | Input On / Continuous | Range | ((V/4000*3) ~ Vmax) V | Resolution | (V/4000) V | Accuracy | 2.50% +/- (V/800) V | Anti-Oscillation | Default/ Osc1/ Osc2/ Osc3/ Disable |
Protection | Over Power (OP) | (Px1.05) +/- (Px0.02) W | Over Voltage (OV) | (Vx1.05) +/- (Vx0.02) V | Over Current (OC) | (Ix1.1) +/- (Ix1.1x0.01/1.05) A | Over Temp (OTP) | 50.00 +/- 5.000 °C | Reverse Max Current (RCP) | (Ix1.1) A | Short Max Current | (Ix1.02) A | Remote Inhibit (RI) | Short | Fault Indicator | SPDT Relay (30Vdc/0.5A or 125Vac/0.25A |
General | AC Input | 95~240 Vac 48~62 Hz | Derating for higher temperatures | (-)1.67% Rated Power / °C | Operating Temperature | 5 °C ~ 40 °C | Fluid | Valve | Normal closed | Flow Rate (Pmax) | >= 1.5~9.0 GPM @ 15 °C Fluid In | Derating for higher temperatures | -3% Rated Power / °C | Pressure | < 80 PSI | Pipe Size | 1/2” NPT Female (<=24kW models, 3/4” for higher power) | Decondensation | Valve thermal control | Transient Mode | Frequency Range | 0.100 - 10,000 Hz | Duty Range | 1.000 - 100.0% | Transient Time Accuracy | 10.0% +/- 50% of Min Time |
Dielectric Strength | Primary Circuit To Chassis | 1500 Vac for 1 min | Primary Circuit To Load Terminal | 1500 Vac for 1 min | Load Terminal To Chassis | 1500 Vdc for 1 min |
|
|
|
|
|
|
|
|
|
|
|
|
|
Descripcion
- La selección más amplia de modelos: 6kW, 9kW, 12kW, 18kW, 24kW, 36kW. Adicional a los modelos estándar, hay disponibles modelos hasta 250kW
- Clasificaciones estándar de voltaje de 60V, 120V, 400V, 600V, 800V y 1000V
- Los modelos PLW ofrecen una huella ultra-compacta y presume una de las mayores densidades de potencia en la industria, 18kW en 2U
- Anti-condensación: Circuito de control de temperatura ingeligente totalmente integrado y válvula de solenoide
- Controladores para LabWindows y LabVIEW y conjunto de comandos SCPI estándar
Las soluciones tradicionales de Carga Electrónica DC son voluminosas. La mayoría son ofrecidas con clasificaciones estándar de voltaje, corriente y potencia. En el mundo ATE, el espacio de rack es un bien altamente codiciado y las demandas de aplicación se encuentran diversificándose constantemente con el desarrollo de nueva tecnología.
La serie PLW de Cargas Electrónicas DC "refrigeradas por agua" de AMREL ofrece a la industria la huella más pequeña, la más alta densidad de potencia y clasificación de corriente, junto con la más amplia selección de modelos de alto voltaje en el mercado. Los modelos PLW son capaces de ser personalizados para cumplir con los requisitos de su aplicación.
Características Clave:
Calibración de caso cerrado Con la línea de eLoad, ya no hay necesidad de enviar su carga electrónica de regreso a la fábrica para ser calibrado, ni de remover decenas de tornillos para llegar a un potenciómetro. Simplemente sigua la rutina de calibración AMREL desde el panel frontal y usted debería estar listo para continuar en un periodo muy corto de tiempo (se requiere algo de equipo de prueba electrónica). Esto virtualmente elimina el tiempo muerto y erradica los costos anuales asociados con enviar su carga electrónica de regreso a la fábrica para ser calibrado.
Protección FET individual Para asegurar la confiabilidad de la serie PLA, el diseño de AMREL incluye protección individual FET. Una carga electrónica programable puede contener muchos FETs en paralelo, que pueden crear una falla en cascada si uno de ellos se pone en corto. El diseño de las cargas electrónicas programables eLoad de AMREL aisla fallas para que otros componentes no se vean afectados o estresados, incrementando el nivel de protección del sistema contra fallas catastróficas. Con protección individual FET, el MTTR (tiempo promedio para reparaciones) se reduce y la carga electrónica regresa rápidamente a operación completa.
Operación de voltaje ultra-bajo El diseño del PLA permite a la carga electrónica programable operar a niveles de voltaje próximos a 1V. Tipicamente disiparán la corriente de clasificación completa por debajo de 1% de su voltaje máximo clasificado. Por ejemplo, una unidad de 60V diseñado para disipar 1500A le permitirá al usuario operar a 0.6V y aún así disipar la cantidad total.
|
|
|
|
|
|
|
|
|